24V Medical Power Supply 200W

600w-kaiguandianyuan

Input Condtion
Model NO.
Output Rail

Output
current

Voltage Accuracy
Regulation
Efficiency
Max Power
Line
Load
90~264V~
47~63Hz
Input current
 4.5A Max.
KRP-G207-U
KRP-G207-UA
KRP-G207-UM
KRP-G207-U5
+12V
16.5A
±5%
±1%
±1%
87%
200W
KRP-G208-U
KRP-G208-UA
KRP-G208-UM
KRP-G208-U5
+15V
12A
±5%
±1%
±1%
87%
200W
KRP-G205-U
KRP-G205-UA
KRP-G205-UM
KRP-G205-U5
+18V
11.1A
±5%
±1%
±1%
87.3%
200W
KRP-G209-U
KRP-G209-UA
KRP-G209-UM
KRP-G20-U5
+24V
8.4A
±5%
±1%
±1%
87.6%
200W
KRP-G20G-U
KRP-G20G-UA
KRP-G20G-UM
KRP-G20G-U5
+28V
 7.2A
±5%
±1%
±1%
88.3%
200W 
KRP-G20J-U
KRP-G20J-UA
KRP-G20J-UM
KRP-G20J-U5
+36V
5.6A
±5%
±1%
±1%
88.3%
200W
KRP-G20T-U
KRP-G20T-UA
KRP-G20T-UM
KRP-G20T-U5
+48V
4.2A
±5%
±1%
±1%
88.3%
200W

24V Medical Power Supply 200W

Features:

  • Design for BF application

  • Safety Class II & EMI Class B

  • 12V/0.5A output for fan

  • High mechanical torque start-up

  • -20℃ to 70℃ operating temperature

  • 5,000m operation altitude

  • Convection cooling for rated load

  • Forced air for max. load

  • U-shape case power supply

General Specifiations:

  • Input voltage: 90V AC to 264V AC

  • Input frequency: 47Hz to 63Hz

  • Inrush current (cold start at 25℃): <30A at 1150V AC or <60A at 230V AC

  • Hold up time: 20ms typical 

  • Over-voltage protection: latch off

  • Short-circuit protection: auto recovery

  • Over load protection: auto recovery

  • Operating temperature: -20℃ to 70℃ derating: 2.5%/℃>50℃ for convection cooling

  • Storage temperature: -40℃ to +85℃

EMI:

  • EN55022 curve "B"

  • EN61000-3-3

  • Harmonics: EN61000-3-2 class D

EMS:

  • EN61000-4-2, -3, -4, -5,-6,-8,-11

Safety:

  • UL/CSA/IEC60950-1,2nd edition

  • ANSI/AMMI/CSA/IEC60601-1,3nd edition

  • CB report,CE mark,RM report/file

Notes:

1. Size:  3.3" (83.8mm) x 5"(127.0mm) x 1.7"(43mm)

2. Mounting Hole:

  • Side edge: 12.5 x 114.3(mm)

  • Bottome: 71.1 x 101.6(mm)

3. Connectors:

  • AC input: Molex 5277-02A or equivalent

  • DC output: Terminal blocks (default for KRP-G207-U) or Molex 5273-08A (default for others) or equivalent

  • Fan, Remote sense: Molex 5045-02A or equivalent

  • Output Pin assignment:1-Vo 2-Vo 3-GND 4-GND

4. Packing:

  • Net weight: 539g approx./ unit

  • Gross weight: 15.3 kg approx./carton, 24 units/carton

  • Carton size: 384 (L) x 339 (W) x 327 (H)

certification-dd